Abstract
An aerosol gas exchange system (AGES) for nanoparticle sampling at elevated temperatures was developed, modeled, and further characterized with laboratory tests with respect to gas exchange efficiency and particle losses. The model describing the gas exchange was first verified with oxygen and later studied with several inert gases having molecular masses between 18 and 135 u. The exchange rate of the lightest compounds exceeds 90% efficiency at the flow rates used. In order to reach similarly high removal efficiencies for larger molecules, the residence time in the AGES has to be increased. The removal of sticky gases was studied with gaseous sulfuric acid. Results agreed with the model where the boundary condition is zero concentration on the wall. The AGES exhibits very limited particle losses (
Original language | English |
---|---|
Journal | Scientific Reports |
Volume | 9 |
Issue number | 1 |
DOIs | |
Publication status | Published - 1 Dec 2019 |
Externally published | Yes |